Electrocardiographs (ECG), electromyographs (EMG), and electroencephalographs (EEG) are medical devices that monitor the electrical activity of the heart, muscles, and brain by measuring the potential differences on the surface of living tissues. These measurements are essential for diagnosing various physiological conditions. However, clinicians often encounter practical challenges when performing biopotential measurements, which can impact the accuracy and reliability of the data collected.
Biopotential electrodes play a crucial role in these measurements. They detect the flow of ions within the body, which is associated with nerve stimulation and muscle contraction. Proper electrode placement and good skin contact are fundamental to obtaining clear and consistent signals. Unfortunately, this step is sometimes overlooked due to time constraints, leading to signal interference or poor data quality.
Variations in patient characteristics, such as age, race, and skin condition, can also affect the impedance levels between the electrodes and the skin. For example, gold electrodes used in ECGs typically have higher impedance compared to silver/silver chloride electrodes commonly found in EMG and EEG systems. In addition, external factors like ablation, electric cautery, defibrillation, and pacemaker telemetry can introduce noise and distort the readings, making accurate measurement more difficult.
To address these challenges, system design plays a key role. Careful planning of the signal conditioning circuitry ensures more reliable and stable biopotential measurements, reducing the need for frequent electrode replacements and saving both time and resources. One important factor is the input bias current of the front-end amplifier. If the skin contact is poor, the input bias current may cause electrode polarization, leading to signal distortion.
Devices such as the AD8625/AD8626/AD8627 family of JFET-input operational amplifiers offer an input bias current of less than 1 pA, while the AD8220 and AD8224 JFET-input instrumentation amplifiers provide even lower bias currents—under 20 pA. These low-current components help maintain signal integrity, especially in environments where high-impedance measurements are required.
Amplifiers that operate over a wide supply voltage range are particularly useful in noisy clinical settings, such as emergency rooms and operating theaters. The AD8625/AD8626/AD8627 series can function from a single 5V to 26V power supply, offering flexibility and stability. Meanwhile, the AD8220 and AD8224 can be powered by either a dual ±18V supply or a single 5V supply, providing rail-to-rail output and maximizing dynamic range. Their low quiescent current of 750 μA makes them ideal for battery-powered applications.
Additionally, the AD8224 can be configured as a single-channel differential output instrumentation amplifier, offering enhanced noise immunity and improved signal resolution. This makes it a versatile choice for a wide range of biopotential measurement systems, ensuring accurate and reliable results in real-world clinical environments.